HomePage > News > Pure Water Conductivity according to the European Pharmacopoeia 

Pure Water Conductivity Measurements according to the European Pharmacopoeia

The purity of water for use in the pharmaceutical industry is mainly controlled by its conductivity.
To comply with the European Pharmacopoeia, this measurement must be made according to chapter 2.2.38.

All low conductivity measurements are influenced by the CO2 content of the air. The CO2 effect contributes significantly to the overall uncertainty of the measurements. Measuring in a closed system is therefore preferable and a slight flow is advantageous.

The 26.6 µS/cm conductivity standard is prepared strictly as described in the Pharmacopoeia and high quality glass bottles are used for storage in order to avoid contamination.

Using the equipment and procedure described below ensures full compliance with chapter 2.2.38 of the European Pharmacopoeia. The control of "water for injections" (aqua ad iniectabilia) and other pure waters is dealt with.

Kit for water conductivity according to the
European Pharmacopoeia
Click here to display
Product Specifications
"Conductivity of Water according to the
European Pharmacopoeia" Application Note



Click here to download
this Free PDF Application Note
Low Conductivity Standard
(26.6 µS/cm at 20°C)
according to the
European Pharmacopoeia



Click here to display
Product Specifications
 
If you are also interested in Pure Water Conductivity Measurement according to USP (US Pharmacopoeia), please follow this link: Conductivity measurement on pure water according to USP24-NF19
 


*
In order to view this PDF format file, you need Adobe Acrobat Reader software version 5.0 or higher. It is available free of charge from the Adobe web site.